3 Function Patient Monitor CMS5100

EGP13,999.00

Performance

3.1 NIBP

Using Oscillometry technology(also called Oscillography technology) to measure the blood pressure.

  • Measurement mode: manual/auto/continuous
  • Measurement interval in auto mode: 1, 2, 3, 4, 5, 5×n(n=2, 3,4…51) min.
  • Resolution: 1 mmHg
  • Accuracy: max mean error: ±5 mmHg; max standard deviation: 8 mmHg.
  • Overpressure protection: dual protection for both software and hardware
  • Others: reset, self-testing and accuracy testing of static pressure
  • Measurement range:

Adult

  1. SYS 40 mmHg~270 mmHg
  2. DIA 10 mmHg~215 mmHg

Pediatric:

  1. SYS 40 mmHg~200 mmHg
  2. DIA 10 mmHg~150 mmHg

Neonate:

  1. SYS 40 mmHg~135 mmHg
  2. DIA 10 mmHg~100 mmHg

3.2 SpO2

Using Photoelectric Oxyhemoglobin Inspection Technology combined with Capacity Pulse Scanning & Recording Technology to measure the SpO2.

  • Measurement range: 0 %~100 %
  • Alarm range: 0 %~100 %
  • Resolution: 1 %
  • Accuracy: 70 %~100 %, ±2 %; below 70 % unspecified.

3.3 PR

Using Photoelectric Oxyhemoglobin Inspection Technology combined with Capacity Pulse Scanning & Recording Technology to measure the pulse rate.

  • Measurement range: 25 bpm~250 bpm
  • Alarm range: 25 bpm~250 bpm
  • Resolution: 1 bpm
  • Accuracy: ±2 bpm or ±2 %, whichever is greater

3.4 TEMP

Using thermal resistor method to measure body surface temperature or body cavity temperature in an accurate, safe and convenient way.

  • Measurement range: 0 ℃~50 ℃
  • Alarm range: 0 ℃~50 ℃
  • Resolution: 0.1 ℃
  • Accuracy: ±0.1 ℃

3.5 Safety and Power supply

  • Power supply: rated voltage: ~220 V, rated power: 50 Hz
  • Safety classification: Class I equipment, type BF defibrillation-proof applied part

Accessories

  •  One User Manual
  • One power cord
  • One power adapter
  • One SpO2 sensor
  • One adult NIBP cuff
  • One NIBP extension tube
  • One TEMP probe

Physical characteristic

  • Dimension: 190 mm(L)×162 mm(W)×240 mm(H)
  • Weight: about 1.6 kg
Reviews (0)
التصنيف:

الوصف

Introduction

  • CMS5100 Patient Monitor is a feature-rich monitoring device that can be applied to adults, pediatric and neonate.
  • User can choose different measurement parameters according to different needs.
  • The patient monitor can monitor NIBP, SpO2, PR and TEMP.
  • It integrates the function of parameter measuring and displaying into a compact, lightweight patient monitor, which is suitable for all levels of hospitals, community medical and home use.

Features

  • Be applicable for NIBP, SpO2 and TEMP monitoring of adult, pediatric and neonate of all ages, easy to operate, high cost-effective
  • Be applicable for medicine, surgery, operating room, ICU/CCU, emergency room, obstetrics and gynecology, pediatrics
  • Compact and flexible appearance, easy for carrying and suitable for indoor and outdoor (in ambulance) monitoring
  • Built-in rechargeable lithium polymer battery, ensures uninterrupted monitoring;
  • Perfect menu design, user-friendly interface
  • Display the measurement result of SYS, DIA, SpO2, PR and bar graph by the display screen of high-brightness digital tube;
  • 2.8” (320×240) true color TFT LCD screen, displays the information of TEMP data, time, SpO2 Plethysmogram, alarm condition, trend graph, list and system settings, etc.
  • Visual and audible alarm for SYS, DIA, MAP, SpO2, PR and TEMP, and upper and lower limit of alarm can be set as necessary
  • Independent nonvolatile memory, storage for up to 2,000 groups of NIBP data, 7900 groups of TEMP data and 78,000 groups of SpO2 data
  • Convenient and quick in reviewing measurement data, available for reviewing the NIBP trend graph of 24 hours, TEMP data of 24 hours and SpO2 trend graph of 22 hours.