Performance
3.1 NIBP
Using Oscillometry technology(also called Oscillography technology) to measure the blood pressure.
- Measurement mode: manual/auto/continuous
- Measurement interval in auto mode: 1, 2, 3, 4, 5, 5×n(n=2, 3,4…51) min.
- Resolution: 1 mmHg
- Accuracy: max mean error: ±5 mmHg; max standard deviation: 8 mmHg.
- Overpressure protection: dual protection for both software and hardware
- Others: reset, self-testing and accuracy testing of static pressure
- Measurement range:
Adult
- SYS 40 mmHg~270 mmHg
- DIA 10 mmHg~215 mmHg
Pediatric:
- SYS 40 mmHg~200 mmHg
- DIA 10 mmHg~150 mmHg
Neonate:
- SYS 40 mmHg~135 mmHg
- DIA 10 mmHg~100 mmHg
3.2 SpO2
Using Photoelectric Oxyhemoglobin Inspection Technology combined with Capacity Pulse Scanning & Recording Technology to measure the SpO2.
- Measurement range: 0 %~100 %
- Alarm range: 0 %~100 %
- Resolution: 1 %
- Accuracy: 70 %~100 %, ±2 %; below 70 % unspecified.
3.3 PR
Using Photoelectric Oxyhemoglobin Inspection Technology combined with Capacity Pulse Scanning & Recording Technology to measure the pulse rate.
- Measurement range: 25 bpm~250 bpm
- Alarm range: 25 bpm~250 bpm
- Resolution: 1 bpm
- Accuracy: ±2 bpm or ±2 %, whichever is greater
3.4 TEMP
Using thermal resistor method to measure body surface temperature or body cavity temperature in an accurate, safe and convenient way.
- Measurement range: 0 ℃~50 ℃
- Alarm range: 0 ℃~50 ℃
- Resolution: 0.1 ℃
- Accuracy: ±0.1 ℃
3.5 Safety and Power supply
- Power supply: rated voltage: ~220 V, rated power: 50 Hz
- Safety classification: Class I equipment, type BF defibrillation-proof applied part
Accessories
- One User Manual
- One power cord
- One power adapter
- One SpO2 sensor
- One adult NIBP cuff
- One NIBP extension tube
- One TEMP probe
Physical characteristic
- Dimension: 190 mm(L)×162 mm(W)×240 mm(H)
- Weight: about 1.6 kg